TO SET UP/EDIT A STEPPED RATE PROFILE:
1
. From the Stepped Definition screen for a Stepped
Rate profile, choose the
Direction button to toggle
the Flow Mode between Infuse and Withdraw
operation.
2. Choose the
Rates button to enter the Set Start
and End Rates screen.
3. Select the
Start Rate button (it should be selected
by default as indicated by the active data entry
box to the right of the button). Enter the flow rate
value desired using the keypad or select the
Minimum Rate or Maximum Rate button.
4. If the Units of Rate are not correct for your
application, choose the Units button.
5. On the Units Selection screen, select the unit of
measure (ml, µl, nl or pl) and the unit of time (sec,
min, hr) as appropriate, then choose Accept to
confirm your selections.
6. Repeat Steps 3 to 5 for the End Rate by selecting
the
End Rate button and continuing data entry.
7. When all parameters are correctly set on the Set
Start and End Rate screen, choose Accept to
confirm your settings and return to Stepped
Definition screen.
8. Choose the
Time and # Steps button to enter
the Set Time & Steps screen.
9
. Select the
T
otal Time
b
utton (it should be selected
by default as indicated by the active data entry
box to the right of the button). Enter the desired
value using the keypad.
10. If the Units of Time are not correct for your
application, select the
Units button to toggle
between sec (seconds) and h:m:s
(hours:minutes:seconds). To use the h:m:s option,
enter your time value in hours, minutes and
seconds, separating each value with a colon.
11. Select the
Number of Steps button and enter the
desired value using the keypad.
12. When all parameters are correctly set on the Set
Time and Steps screen, choose Accept to confirm
your settings and return to the Stepped Definition
screen.
Set Start and End Rate screen