Manual
KDS Allegro Peristaltic Pump System User’s Manual
phone 508.429.6809 fax 508.893.0160
e-mail info@kdscientific.com web www.kdscientific.com
KD Scientific Allegro Peristaltic Pump System User’s Manual
Publication 5613-004-REV-A
33
Operating the KDS Allegro Peristaltic Pump System
Select Program
Review Program Steps
(Pump Operation Profiles)
S
elect Tubing & Size
OR
Create New Tubing Size
Enter Pump Operation Profile
Parameters
Validate Commands and
I/O Operations
C
ustom Program Workflow
The following flow diagram describes the use of
a Custom Program:
The software provides the user with a variety of prompts
and warnings to help guide the setup of programs. For
example, if a pump Program incorporates steps that
exceed the capacity of the defined tubing, the system will
prompt you that parameters are out-of-range. The system
also prompts users if required values are not entered.
✐
NOTE: The pump will check the validity of
parameters entered for each Step in the
Program, but will not review the Program in
its entirety to verify that it is logical and can
be fully executed