Configuring the KDS Allegro Peristaltic Pump System
KDS Allegro Peristaltic Pump System User’s Manual
phone 508.429.6809 fax 508.893.0160
e-mail info@kdscientific.com web www.kdscientific.com
KD Scientific Allego Peristaltic Pump System User’s Manual
Publication 5613-004-REV-A
Tube Run-In
New tubing or tubing that has been idle for extended
periods should be run-in prior to use. The pump comes
standard with programs to complete this task. Any tube
s
ize that meets the diameter specifications for the
appropriate pump head can be used.
1. From the Mode Selection screen, select the
proper program for the head configuration of
the pump.
Please select program
I_BREAK_IN for the
Allegro I, II_BREAK_IN for the Allegro II or
III_BREAK_IN for the Allegro III.
2. Once the proper tube break-in program is
selected, the Program Configuration screen
will open.
3. Select the Run button to start the break-in
program. The program run time is 15 minutes.
Tubing Break-In Selections
Select Run to start the
break-in program
✐
NOTE: Large diameter and/or high
durometer tubing require lubrication of
tubing. See Appendix E for ordering
information.