User guide

2
KDS Model 310 Plus Series
Specifications
Model Model 310 Plus
Syringe Size 1–250 microliter
Electrical Rating US model 115V
~
, 0.1A
CE model 230V
~
, 0.06A
Fuse 5x20 mm, 250V
~
slow blow, 0.1A
Voltage Operating Range US Model 100-120V
~
, 50/60Hz
CE Model 200-240V
~
,50/60Hz
Drive Mechanism Microprocessor controlled stepper motor
(
1
2 microstep), leadscrew
Force 2 lbs (minimum)
Pusher Advance Per Half Step 1.58 microns
Minimum Stepping Rate One
1
2 step every 4 seconds
Maximum Stepping Rate 916
1
2 steps/second
Flowrate Range 0.001–363.7 µl/min [0.5–250 µl syringes]
Dimensions 9x6x3 inch (23x15.25x8 cm) controller
7x1.7x2 inch (17.8x4.4x5.1 cm) injector
Weight 4.5 lb (2 Kg)
Atmospheric Specifications
Temperature 4˚C to 40˚C (40˚F to 104˚F)
Humidity 20% to 80% RH, non-condensing
Mode of Operation Continuous
Classification Class I
Pollution Degree 2
Installation Category II
Output N/A
Physiological Effects N/A
Cooling Conditions No special considerations
Mechanical Stability No special considerations
Protective Packaging No special considerations
Earth Terminals No External connections required
Removable Protective Means N/A
Supplier Name KD Scientific Inc.
Address 84 October Hill Road, Holliston, MA 01746