Chapter 11: Qualification Reports
Kaye ValProbe RT User Manual 102
11.2. Qualification Detailed Report
Detailed Reports are created from the raw encrypted Qualification files and provide the
complete details of all sensor measurements and calculations (Statistical, Lethality, Steam
Saturation) during a Qualification study.
The content of the Qualification file is based on the parameters defined in the Setup. The
Setup defines the number and type of sensors utilized, the arrangement of sensors into
Groups, the specific calculations to be performed, the Start and Stop conditions for
Qualification and Exposure, and the data storage rate.
Prior to generating the Detailed report, you have the option to set a separate print rate which
can be a multiple greater than the storage rate; allowing you to maximize the accuracy for
graphs and calculations (i.e. storing data every ten seconds but printing once a minute) and
minimizing the number of pages printed.
The main report header provides traceability information for the study including:
Company name
Date printed, username
Study name
Created by
Created date / time
Study started by with date and time
Asset name
Asset ID
Serial number
User comments.
Reporting Software
Firmware and Software versions
The remaining sections of the cover page include, the groups included in the report, any
sensors excluded from group calculations, the sensors used in the study by serial number,
and any user comments added during post-qualification reporting.
The body of the report is arranged by Groups. For each Group the Detailed report contains
pages detailing the sensor measurements, the Statistical calculations, Lethality calculations (if
defined), and the Saturation Temp and Pressure calculations (if defined).The last pages of the
report list the System and Event messages along with a Footer as defined in report options.
Figure 116: Qualification Detailed Report