Chapter 10: Sensor Verification
Kaye ValProbe RT User Manual 92
Choose the Temperature verification type on the Select Verification Type window.
Figure 99: Select Verification Type
On the Temperature Verification Setup screen, the following options are available:
Choose between one to six temperature setpoints
Define for each setpoint the setpoint temperature, Pass/Fail criteria and results log
times for reports
Define sensor stability and minimum duration
IRTD temperature reference stability criteria and duration
Additional time at stability to ensure sensors, loggers, and references are stable
SOP protocol numbers or other information
Figure 100: Setup Temperature Verification
To enter additional Setpoints or delete setpoints click on the appropriate icon. Once all
setpoints and criteria are entered, save the Verification Setup under a specific name. If
desired, setup printing is available by selecting the Printer icon. To delete a setup, click the
Trash Bin icon.
Multiple Setups can be generated and saved with different configurations, all are listed on the
right of the screen by Setup name and sorted by creation date. To modify an existing setup,
double-click on it to open it, modify it and then click Save to save the changes or Cancel to
revert your changes.