Users Manual Part 2

Table Of Contents
Chapter 11: Qualification Reports
Kaye ValProbe RT User Manual 132
Figure 149: Report Options Screen
Press Generate Report to generate the respective report type or customize report options.
Content for the selected report can be customized via the Report Content button. A table
opens, listing all available groups and valid calculations for the sensors type of the group.
Lethality and MKT calculations cannot be selected for one single report. Saturation,
temperature, and saturation pressure calculations can be selected or deselected for all groups
together in a report.
Availability of a saturation calculation is dependent on whether there is a pressure sensor and
temperature sensor defined in the study. The calculation is selectable per group for all
sensors. If the calculation for all sensors is selected, group selection is automatically
overridden.
Figure 150: Report Content Selection - Detailed Report
Inside the Summary Report, you can choose between Lethality and MKT calculations. In
addition, there are options to include a Setup Report (including the wiring diagram if specified)
and a study overview graph.