Users Manual Part 2

Table Of Contents
Chapter 11: Qualification Reports
Kaye ValProbe RT User Manual 122
Figure 130: Mark Cycles Using Events
If creating a new event, the user can specify a message which appears in the Detailed Report
and in the graph. The user can also specify "When" which defines when the event occurs, as
well as a “Label” and the condition for the event. (e.g. min temp > 100.0 C). When all details
are entered press the Checkmark and the selected time will be marked
When completed with any edits to the Qual / Exposure times press Define Additional Cycles.
Figure 131: Define Additional Cycles screen
The Define Add’L Cycles screen can be utilized if additional cycles beside Qual and
Exposure need to be defined. If no additional cycles need to be defined simply press Report
Parameters at the top of the screen to proceed to the next step. The system allows up to 15
additional cycles to be defined. Press the button at the top right of the screen to go to
Cycles window. The method and functionality for defining cycles is the same as defined in
Exposure Start / End.
In the Cycles window all created cycles are displayed with the defined Start and End times.
The Cycles window also allows modifications of cycles names. UnMark Cycle can be used to
delete a cycle.