User's Manual

d. Long drip timing: set the running time of the pump during calibration; Before proceeding to the next step, make sure that the pump pipe inlet
has been immersed in water, the pump pipe outlet is placed in the measuring tube;
e. Start titration: click the titration button and the pump will stop after running the duration of the previous set;
f. Input volume: after reading the volume of liquid in the barrel input, in ml;
g. Calibration complete: Click the "Calibration complete" button to complete the flow calibration.
4.6 Firmware Upgrades
When the pump's firmware program is updated, the user needs to upgrade the firmware to use.
The upgrade steps are as follows: Enter the App setting interface, if you find a new version of the firmware appears, click the update button to update
the firmware. At this time, do not perform other operations, do not exit the App or re-enter the App. Wait until the upgrade is complete, and the device
screen restarts after the upgrade is complete. After the device upgrade is completed, normal operations can be performed. If the upgrade fails, please
repeat the upgrade steps.
Note: You cannot power down during the upgrade and the app does not take any other action during the upgrade process.
5 Appendix
5.1 Technical Parameters
Dimension(LxWxH) 75x70x110 mm
weight 534gDoes not include a power adapter
power adapter
Input100VAC -240VAC
outputDC24V 1.9A
Dosing paramete r
Pump Head:3 Rotor KPAS100
Flow<=110ml/min
Precision<±2%
inte rface Wi-Fi Rotating Encoder/Wi-Fi
working e nvironment Temperature 0 - 70 ° C, Humidity 10% - 90% (non-condensing)
Storage environment Temperature-20 - 85 ,Humidity 10% - 90% (non-condensing)
5.2 After-sales warranty information
1. Warranty conditions The free service during the warranty period is valid only for normal use and maintenance according to the user manual.
Any malfunction or damage caused by human beings is not covered by the warranty. Users should take good care of the purchase invoice and user
manual so that you can get satisfactory after-sales service in time.
2. Warranty scope The company will provide free warranty service for any damage caused by manufacturing processes or components within one
year from the date of purchase. The free repair service provided during the warranty period includes free repair, free replacement and replacement
of faulty spare parts, and products that cannot be repaired are replaced by the same model (the model has been discontinued, and the model is
similar). The free service does not include shipping costs for the product due to repairs.
3. Non-warranty scope The following factors are not covered by the free warranty, and customer repairs are subject to a fee.
1) Product appearance (please confirm at the time of purchase);
2) Improper use, maintenance or storage (please use, maintain and keep it in accordance with the user manual);
3) Access to an improper power source;
4) Damage to components caused by short-circuiting of the circuit board caused by various types of insects entering the machine;
5) Losses due to accidents;
6) Use of inappropriate spare parts (not applicable to our spare parts);
7) Those who are not authorized by the company are negligent in handling, modification or repair (please do not disassemble and repair);
8) malfunction or damage caused by use outside the applicable place;
9) Damage caused by force majeure;
10) consumable parts (such as pH electrodes, ORP electrodes, etc.);
11) The warranty period has expired.
Kamoer, Kamoer text and icons are registered trademarks of Kamoer Fluid Tech (Shanghai) Co., Ltd. The company reserves the right to make
improvements and changes to the product's appearance and technical specifications without prior notice.
6 update log
date detail
2021-07-27 Write an online user manual
2022-03-11 Add fcc warning