User's Manual
a. Device information:Display basic information of the device
b. Firmware version:Display the current version of the firmware information;
c. Device name: Device name, the user can modify it as needed;
d. Serial number:equipment serial number;
e. Firmware update :If the device has new firmware, click here to update;
f. Pump tube setting:Set the life of the pump tube and check the use time of the pump tube here;
g. Connect to the smart controlle r:Click to enter the QR code for connecting to the smart controller, and connect through the smart controller in
the Aqua Cloud (see the introduction of the smart controller for specific operations)
h. Flow calibration: Perform flow calibration of the device here;
i. Restore factory settings:After clicking, the device parameters will be restored to the factory default settings;
j. Delete device:Click to unbind the app and device.
4.5 Flow Calibration
The purpose of calibration is to improve the accuracy of the added elements;
Click "Flow Calibration" in the setting interface to enter the calibration interface.
Calibration needs to use the barrel, the pump factory has been equipped with a 10ML measuring tube, taking into account the pump tube titration
concentration is different, the degree of aging of the pump tube is different, the first use needs to be calibrated, suspected titration is not accurate when
it is timely calibration.
a. Flow rate calibration: click on the flow rate calibration in the setting s --coma interface, and bring out the flow rate calibration channel selection
interface;
b. Speed setting: set the speed of calibration;
c. Start emptying: the purpose of emptying is to let the air in the pump tube drain, so that the accuracy of the rear calibration will not be affected;