1. Tube type: Click to enter the tube type calibration interface;
2.Start priming: Click to start Priming, click again to stop priming, when calibrating, you need to
empty the air inside the pump tube;
3. Flow rate and speed: the flow rate before calibration and the speed of the corresponding tube
type;
4. Long dosing: The length of time the calibration needs to be run. The longer the calibration time
is set, the calibration will be more accurate if the calibration container allows it.
5. Start dosing: Click to start dosing, the pump starts running, when the calibration time
countdown is completed, the pump stops running, input the measured volume, click calibration to
complete the calibration;
6. Measured volume: the volume obtained when the calibration pump is finished running;
4MAINTENANCE
4.1 Chemical compatibility
The standard configuration of the pump tubing is a silicone tubing that can be used in most
applications. If you are delivering a specific solution, please contact the factory.
Contact the manufacturer to determine chemical compatibility and match the appropriate pump
tubing.