User's Manual

a.Current VersionDisplay the current version of the firmware, if the firmware is updated, there will be a prompt below;
b.Name The name of the device can be modified here to identify the purpose of the device;
c.UpdateFirmware update, if new firmware is released, there will be a prompt;
d.Tube se tting Set the life of the equipment tube here and check the length of time the equipment tube is used;
e.Flow CalibrationPerform the flow calibration of the device here, see the next section for details: flow calibration;
f.Factory Setting RestAfter clicking, the device parameters will be restored to the factory default settings;
g.Remove Device Click Unbind App and device binding.
4.6 Flow calibration
The purpose of calibration is to improve the accuracy of equipment operation; In the setting interface, click "flow rate calibration" to enter the flow
calibration interface. Calibration requires the use of a measuring cylinder. The device is equipped with a 10ML measuring cylinder at the factory.
Considering the different concentration of the titration solution of the equipment tube and the degree of aging of the equipment tube, calibration is
required for the first use. When the titration is inaccurate, it should be calibrated in time.
a. Flow Calibrationclick flow rate calibration on the setting interface to call up the flow rate calibration channel selection interface;
b. Rotation set the calibrated speed;
c. Start emptyingThe purpose of evacuation is to let the air out of the equipment tube, so that the accuracy will not be affected during the