OperationManual

- 9 -
ENGLISH
Guidance and manufacturer’s declaration – electromagnetic immunity
The model CL-R813 is intended for use in the electromagnetic immunity environment (Home
healthcare environment dened in IEC 60601-1-2) specied below.
The customer or the user of the model CL-R813 should assure that it is used in such an
environment.
Immunity
IEC 60601
test level
Compliance
level
Electromagnetic
environment –guidance
Proximity
electromagnetic
eld from RF
wireless
communication
equipment
(Specied the table
below)
Portable and mobile RF
communications equipment
should be used no closer to any
part of the model CL-R813,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and 80 MHz
80 % AM at 1 kHz
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and 80 MHz
80 % AM at 1 kHz
Recommended separation
distance
d
= 1.2
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
d
= 1.2 80 MHz to 800 MHz
d
= 2.3 800 MHz to 2.7 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and
d
is
the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol: