Žæˆµà–¾‘
- 8 -
English
Guidance and manufacturer’s declaration – electromagnetic immunity
The model CL-R190 is intended for use in the electromagnetic immunity environment (Home
healthcare environment dened in IEC60601-1-2) specied below. The customer or the user of
the model CL-R190 should assure that it is used in such an environment.
Immunity IEC 60601 test
level
Compliance
level
Electromagnetic environment
– guidance
Proximity
electromagnetic
eld from RF
wireless
communication
equipment
(Specied the table
below)
Portable and mobile RF
communications equipment should
be used no closer to any part of
the model CL-R190, including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and
80 MHz
80% AM at 1kHz
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and
80 MHz
80% AM at 1kHz
Recommended separation
distance
d = 1.2
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz
10 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be less than the
compliance level in each frequency
range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol: