Guidance and manufacturers declaration – electromagnetic immunity
The model CL-R190 is intended for use in the electromagnetic immunity environment (Home
healthcare environment dened in IEC60601-1-2) specied below. The customer or the user of
the model CL-R190 should assure that it is used in such an environment.
Immunity IEC 60601 test
level
Compliance
level
Electromagnetic environment
– guidance
Proximity
electromagnetic
eld from RF
wireless
communication
equipment
(Specied the table
below)
Portable and mobile RF
communications equipment should
be used no closer to any part of
the model CL-R190, including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and
80 MHz
80% AM at 1kHz
3 V
0.15 MHz to 80 MHz
6 V
ISM bands and
amateur radio
between
150 kHz and
80 MHz
80% AM at 1kHz
Recommended separation
distance
d = 1.2
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz
10 V/m
80 MHz to 2.7 GHz
80% AM at 1kHz
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2.7 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be less than the
compliance level in each frequency
range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol: