Thermocouples
Resistance transmitter and resistor/potentiometer
Direct current
Designation Type Standard ITS Measuring range
in °C
Measuring accuracy
a
a
The accuracy values refer to the complete measuring range.
Min. Max.
Pt13Rh-Pt R IEC 584-1 ITS-90 -50 1768 ± 0.15 % from +50 °C
Pt10Rh-Pt S IEC 584-1 ITS-90 -50 1768 ± 0.15 % from +20 °C
Pt30Rh-Pt6Rh B IEC 584-1 ITS-90 0 1820 ± 0.15 % from +400 °C
Fe-CuNi J IEC 584-1 ITS-90 -210 1200 ± 0.1 % from -100 °C
Cu-CuNi T IEC 584-1 ITS-90 -270 400 ± 0.1 % from -150 °C
NiCr-CuNi E IEC 584-1 ITS-90 -270 1000 ± 0.1 % from -80 °C
NiCr-Ni K IEC 584-1 ITS-90 -270 1372 ± 0.1 % from -80 °C
NiCrSi-NiSi N IEC 584-1 ITS-90 -270 1300 ± 0.1 % from -80 °C
Fe-CuNi L DIN 43710 IPTS-68 -200 900 ± 0.1 %
Cu-CuNi U DIN 43710 IPTS-68 -200 600 ± 0.1 % from -100 °C
Chromel-Copel (Ni9.5Cr-Cu44Ni) L GOST R 8.585-2001 ITS-90 -200 800 ± 0.1 % from -80 °C
Chromel-Alumel GOST R 8.585-2001 ITS-90 -270 1372 ± 0.1 % from -80 °C
W5Re-W20Re A1 GOST R 8.585-2001 ITS-90 0 2500 ± 0.15 %
W5Re-W26Re C ASTM E230/E230M-11 ITS-90 0 2315 ± 0.15 %
W3Re-W25Re D ASTM E1751/E1751M-09 ITS-90 0 2315 ± 0.25 %
PL II (Platinel
b
II)
b
Platinel is a registered trademark of Engelhardt Corp.
ASTM E1751/E1751M-09 ITS-90 0 1395 ± 0.15 %
Cold junction Pt1000 internal or external cold junction; temperature adjustable 0 to 80 °C
Cold junction accuracy ± 1 K
Designation Measuring range Measuring accuracy
Resistance transmitter Up to 10000 Ω ±10 Ω
Resistor/potentiometer ≤ 400 Ω±400 mΩ
≥ 400 Ω to ≤ 4000 Ω± 4 Ω
> 4000 Ω to ≤ 10000 Ω±10 Ω
Connection type Resistance transmitter: 3-wire circuit
(A=Start,S=Slider,E=End)
Resistor/potentiometer: 2-wire circuit, 3-wire circuit, and 4-wire circuit
Sensor lead resistance ≤ 11 Ω per line for 2-wire circuit, 3-wire circuit, and 4-wire circuit
Designation Measuring range Accuracy
a
a
The accuracy value refers to the complete measuring range.
Input resistance
Input for mV generator -100 to 1100 mV ±0.05 % R
E
≥ 1MΩ