Technical data
Analog input
All analog inputs are equipped with a digital filter of second order (filter constant adjustable from 0 to 10 s) and have a sampling rate of > 2 mea-
surements per second.
RTD temperature probe
Designation Standard ITS Connection type Measuring range
in °C
Measuring accu-
racy
a
a
The accuracy value refers to the complete measuring range.
Min. Max.
Pt100 IEC 60751:2008 ITS-90 2/3-wire -100 200 ±0.2 K
Pt500 2/3-wire -200 850 ±0.4 K
Pt1000 4-wire -100 200 ±0.1 K
T
K
= 3.85 × 10
-3
1/K 4-wire -200 850 ±0.2 K
Pt100 GOST 6651-2009 A.2 ITS-90 2/3-wire -100 200 ±0.2 K
T
K
=3.917×10
-3
1/K 2/3-wire -200 850 ±0.4 K
4-wire -100 200 ±0.15 K
4-wire -200 850 ±0.25 K
Pt50 2/3-wire -200 850 ±0.5 K
T
K
=3.91×10
-3
1/K 4-wire -200 850 ±0.3 K
Ni100 DIN 43760 IPTS-68 2/3-wire -60 250 ±0.4 K
T
K
=6.18×10
-3
1/K 4-wire -60 250 ±0.2 K
Ni500 2/3-wire -60 250 ±0.4 K
T
K
=6.18 ×10
-3
1/K 4-wire -60 250 ±0.2 K
Ni1000 2/3-wire -60 250 ±0.4 K
T
K
=6.18×10
-3
1/K 4-wire -60 250 ±0.2 K
Ni 100 GOST 6651-2009 A.5 ITS-90 2/3-wire -60 180 ±0.4 K
T
K
=6.17×10
-3
1/K 4-wire -60 180 ±0.2 K
Cu50 GOST 6651-2009 A.3 ITS-90 2/3-wire -180 200 ±0.5 K
T
K
=4.28×10
-3
1/K 4-wire -180 200 ±0.3 K
Cu100 2/3-wire -180 200 ±0.4 K
T
K
= 4.28 × 10
-3
1/K 4-wire -180 200 ±0.2 K
Connection type 2-wire, 3-wire, or 4-wire circuit
Sensor lead resistance
- For 3/4-wire circuit 11 Ω per line
- For 2-wire circuit Measuring resistance + 22 Ω inner line resistance
Sensor current < 0.3 mA