Datasheet
Page 1/10
JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
e-mail: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax: +44 1279 635262
e-mail: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO PROCESS CONTROL INC.
885 Fox Chase, Suite 103
Coatesville PA 19320, USA
Phone: 610-380-8002
1-800-554-JUMO
Fax: 610-380-8009
e-mail: info@JumoUSA.com
Internet: www.JumoUSA.com
dTRANS T03 T
Type 707032/...
dTRANS T03 TU
Type 707034/...
dTRANS T03 BU
Type 707033/...
dTRANS T03 J
Type 707030/...
dTRANS T03 B
Type 707031/...
V1.01/EN/00391028
JUMO dTRANS T03 J, B, T
Analog 2-wire transmitter
with digital adjustment
JUMO dTRANS T03 BU, TU
Analog 3-wire transmitter
with digital adjustment
for connection to Pt100 resistance thermometers
for installation in: - terminal head Form B to DIN EN 50446
- terminal head Form J
for mounting on: - mounting rail according to DIN EN 60715
Brief description
These transmitters are designed for industrial applications and are used to measure the
temperature through Pt100 resistance thermometers in 2-wire or 3-wire circuit
connections (Pt500 or Pt1000 linearization upon request).
The 4to20mA (2-wire transmitter) or 0to10V (3-wire transmitter) output signal is linear
with temperature.
The continuous analog signal path enables an extremely fast reaction time of the output
to a change in temperature (continuous analog measurement instead of digital sampling
rate), resulting in a low-noise output signal that is insensitive to interference. A very high
degree of precision – even with small ranges – is ensured thanks to the range-specific gain
adjustment.
Digital communication allows the transmitter to be adapted to the measurement task
(range, probe break and fine calibration).
Two versions are available to suit specific requirements:
Instruments with basic type extension 880/990 (adjustable)
The transmitters are calibrated for a fixed range but can, at any time, be calibrated for a
different range through the setup program.
Instruments with basic type extension 881/991 (configurable)
The required range can be configured through the setup program, without sensor
simulation and measurement.
Overview of function
dTRANS T03 J
Type 707030/...
dTRANS T03 B
Type 707031/...
dTRANS T03 T
Type 707032/...
dTRANS T03 BU
Type 707033/...
dTRANS T03 TU
Type 707034/...
Input Pt100 Pt100 Pt100 Pt100 Pt100
Connection
circuit (sensor)
2-wire 2-wire or 3-wire 2-wire or 3-wire 2-wire or 3-wire 2-wire or 3-wire
Mounting terminal head
Form J
terminal head
Form B
mounting rail terminal head
Form B
mounting rail
Output 4 to 20mA 4 to 20mA 4 to 20mA 0 to 10V 0 to 10V
Connection
circuit (output)
2-wire 2-wire 2-wire 3-wire 3-wire
Data Sheet 707030
70703000T10Z001K000