Note:
The sensor technology specified in Page 10 does not have zone separation.
The type of zone separation as well as the cable selection must be implemented or selected in such a way that the defined zone classi-
fications and their requirements continue to be in place.
Use of a probe with EPL "Gb" with a separation element (DIN EN 60079-26). The figure shows a probe with active zone separation ac-
cording to DIN EN 60079-26. Mounting of the terminal head in zone 0 is not permitted.
However, use in zone 0 is permitted below the separation element. The same requirements as in the above figure apply for the zone
classification.
Type 701155/...044
None Ex-area
Ex-area
Zone 0 or 1
Zone 1 or 2
Here an adequate zone separation
is needed that an existing zone definition
is not influenced e.g. through distance or
a cable feed-through that has got
an ATEX-type approval.
RTD temperature probe
When mounting in zone 0,
a suitable impact protection
for the terminal head
is required.
Sensor 1
Sensor 2