40505200T10Z001K000
JUMO DELOS SI
Precision pressure transmitter with switching contacts and display
Application
Food & pharma
CIP/SIP systems
Machine and system construction
Air conditioning and refrigeration system construction
Brief description
The JUMO DELOS SI high-precision, electronic pressure switch with analog output meets tech-
nical requirements relevant to hygiene. In a readily cleanable design and with a front-flush, seal-
free measuring system, it is specifically geared to hygienic process safety standards. The nom-
inal measuring ranges set at the factory extend from 400 mbar to 60 bar. They can be rescaled
in the ratio 1:4, allowing the user to perform several measurement tasks with just one measuring
instrument.
The following output variants are available, depending on the application: PNP or PNP
switching outputs or 1× PNP switching output and an analog output signal.
The process pressure is visualized by a large, positively lit LC display.
Configuration can simply be performed on the measuring instrument, or by using the more con-
venient setup program. The enclosure can be rotated ±160° to make it easier to read, and the
display can be rotated 180° for overhead mounting.
A high-temperature variant is available for hot media up to 200 °C (CIP and SIP cleaning sys-
tems).
Type 405052
Customer benefits
Process safety
An hygienic design, with parts in contact with the medium made from stainless steel. The
sys-tem has no dead zones, which prevents microbial contamination and ensures maximum
pro-cess safety for hygienic processes.
Uncomplicated and time-saving
Save time with easy, local measuring instrument configuration or configuration via a more
con-venient setup program.
Flexible
The enclosure can be turned ±160° and the display rotated 180° for user-friendly and indi-
vidual alignment of the measuring instrument for optimum readability, configuration and
elec-trical installation in any mounting position, as well as for running an auto-zero function.
Special features
Linearity from 0.1 % of MSP (measuring
span)
Scaling of measuring range 1:4
Large, brilliant two-color display with
alarm function
Time-saving and simple configuration on
the instrument or via a more convenient
setup program
Flexible mounting variance by rotating
the display (180°), followed by auto zero-
ing
Enclosure can be turned (±160°) to im-
prove readability
Option for increased medium tempera-
ture up to 200 °C (SIP and CIP capability)

Summary of content (12 pages)