Datasheet

V2.00/EN/00541512
Page 8/18Data Sheet 202552
JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone: +49 661 6003-0
Fax: +49 661 6003-607
Email: mail@jumo.net
Internet: www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM 20 2DY, UK
Phone: +44 1279 63 55 33
Fax: +44 1279 62 50 29
Email: sales@jumo.co.uk
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
6733 Myers Road
East Syracuse, NY 13057, USA
Phone: +1 315 437 5866
Fax: +1 315 437 5860
Email: info.us@jumo.net
Internet: www.jumousa.com
20255200T10Z002K000
Temperature compensation
Measuring circuit monitoring
Two-electrode systems
Four-electrode systems
Binary input
Controller
Type of compensation Range
a
a
Note the sensor operating temperature range!
Linear 0 to 8 %/K -10 to +160 °C
ASTM D1125 - 95 (ultra-pure water) 0 to 100 °C
Natural waters (ISO 7888) 0 to 36 °C
Reference temperature
Adjustable from 15 to 30 °C;
preset to 25 °C (default)
Inputs Underrange/overrange Short circuit Broken lead
Conductivity Yes Depends on measuring range Depends on measuring range
Temperature Yes Yes Yes
Voltage 2 to 10 V
0to10V
Yes
Yes
Yes
No
Yes
No
Current 4 to 20 mA
0to20mA
Yes
Yes
Yes
No
Yes
No
Resistance transmitter No No Yes
Cell constant
[1/cm]
Setting range
of the relative cell constant
Resulting usable range
[1/cm]
0.01
20 to 500 %
0.002 to 0.05
0.1 0.02 to 0.5
1.0 0.2 to 5
3.0 0.6 to 15
10.0 2.0 to 50
Cell constant
[1/cm]
Setting range
of the relative cell constant
Resulting usable range
[1/cm]
0.5
20 to 150 %
0.1 to 0.75
1.0 0.2 to 1.5
Activation Floating contact is open: function is not active
Floating contact is closed: function is active
Function Key lock, manual mode, HOLD, HOLD inverse, alarm suppression, freeze measured value,
level lock, reset day counter, reset total counter, parameter set changeover, flow-rate
measurement
Pulse input for flow measurement Binary input 1: approx. 3 to 2000 Hz, resolution 2 Hz
Binary input 2: approx. 4 to 300 Hz, resolution 0,5 Hz
At the device only one binary input for flow measurement can be used.
Controller type Limit comparators, limit controllers, pulse length controllers, pulse frequency
controllers, modulating controllers, continuous controllers
Controller structure P/PI/PD/PID