Technical data
Analog inputs
RTD temperature probe
Thermocouples
Direct current
Analog output
Designation Measuring range Accuracy
2/3-wire circuit
1
Ambient temperature error
Pt 100 DIN EN 60751 -200 to +850 °C 0.5% / 0.1% 50 ppm
Pt 1000 DIN EN 60751 -200 to +850 °C 0.5% / 0.1% 50 ppm
Connection type 2-wire, 3-wire circuit Maximum output resistance 30
Measuring rate 210 ms
Error tolerance time 5s time taken into account for all diagnosis tests
Input filter digital filter, 2nd priority; filter constant can be set from 0 to 100 s
Special features Individual probe Pt100 2-wire,
display can also be programmed in °F
Designation Measuring range Accuracy
1
Ambient temperature error
Fe-CuNi „L“ DIN 43710 -200 to +900 °C 0.4% 100 ppm
Fe-CuNi „J“ DIN EN 60584 -200 to +1200 °C 0.4% 100 ppm
Cu-CuNi „U“ DIN 43710 -200 to +600 °C 0.4% 100 ppm
Cu-CuNi „T“ DIN EN 60584 -200 to +400 °C 0.4% 100 ppm
NiCr-Ni „K“ DIN EN 60584 -200 to +1372 °C 0.4% 100 ppm
NiCrSi-NiSi „N“ DIN EN 60584 -100 to +1300 °C 0.4% 100 ppm
Pt10Rh-Pt „S“ DIN EN 60584 0 to +1768 °C 0.4% 100 ppm
Pt13Rh-Pt „R“ DIN EN 60584 0 to +1768 °C 0.4% 100 ppm
Pt30Rh-Pt6Rh „B“DIN EN 60584 300 to 1820 °C 0.4% 100 ppm
W3Re-W25Re „D“ 0 ... 2495°C 0.4% 100 ppm
Cold junction Pt 100 internal
Cold junction accuracy ± 1 K
Measuring rate 210 ms
Error tolerance time 5s time taken into account for all diagnosis tests
Input filter digital filter, 2st priority; filter constant can be set from 0 to 100 s
1. The accuracy values refer to the maximum measuring range.
Measuring range Accuracy Ambient temperature error
4 to 20 mA, voltage drop < 2 V 0.2% 150 ppm
Scaling can be freely programmed within the limits
Measuring rate 210 ms
Error tolerance time 5s time taken into account for all diagnosis tests
Input filter digital filter, 2st priority; filter constant can be set from 0 to 100 s
Special features Individual probe 4 to 20mA
Type of signal Accuracy Residual ripple Load influence Temperature error Load resistance
Current 4 ... 20 mA
0.5 % ± 0.5 % bei 300 ± 0.05 mA/100 150 ppm / °C 500
0 ... 20 mA
Voltage 2 ... 10 V
0.5 % ± 0.5 % ± 15 mV 150 ppm / °C 500
0 ... 10 V