User guide

ME
9
2.4. Technical specifications
F25-ME
Working temperature range °C -28 ... 200
Temperature stability °C ±0,01
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.26 0.2 0.06
Refrigerant R134a
Overall dimensions (WxDxH) cm 23x42x61
Bath opening (WxL) cm 12x14
Bath depth cm 14
Filling volume from ... to liters 3 ... 4,5
Weight kg 31
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50
Current draw (at 230 V) A 12
Current draw (at 230 V) CH A <9 + 2>
Mains power connection 230 V/60 Hz V/ Hz 207-253 / 60
Current draw (at 230 V) A 12
Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60
Current draw (at 115 V) A 13
Mains power connection 100 V/50-60 Hz V/ Hz 90-115 / 50-60
Current draw (at 100 V) A 13
F26-ME F32-ME
Working temperature range °C -28 ... 200 -30 ... 200
Temperature stability °C ±0,01 ±0,01
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20
0.26 0.2 0.06
+20 0 -20 -30
0.45 0.39 0.15 0.05
Refrigerant R134a R134a
Overall dimensions (WxDxH) cm 42x42x42 31x42x64
Bath opening (WxL) cm 12x14 18x12
Bath depth cm 14 15
Filling volume from ... to liters 3 ... 4,5 5,5 ... 8
Weight kg 31 37
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current draw (at 230 V) A 12 12
Mains power connection 230 V/60 Hz V/ Hz ------ 207-253 / 60
Current draw (at 230 V) A ------ 12
Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60 103-127 / 60
Current draw (at 115 V) A 13 14
Mains power connection 100 V/50-60 Hz V/ Hz ----- 90-110 / 50-60
Current draw (at 100 V) A ----- 14
All measurements have been carried out at:
rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.