User guide
Operator responsibility – Safety recommendations
10
F33-ME F34-ME
Working temperature range °C -30 ... 200 -30 ... 150
Temperature stability °C ±0,01 ±0,01
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20 -30
0.5 0.32 0.12 0.03
+20 0 -20 -30
0.45 0.32 0.14 0.3
Refrigerant R134a R134a
Overall dimensions (WxDxH) cm 36x46x69 38x58x62
Bath opening (WxL) cm 23x14 24x30
Bath depth cm 20 15
Filling volume from ... to liters 12 ... 16 14 ... 20
Weight kg 44 42
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current draw (at 230 V) A 12 12
Mains power connection 230 V/60 Hz V/ Hz 207-253 / 50 207-253 / 60
Current draw (at 230 V) A 12 13
Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60 103-127 / 60
Current draw (at 115 V) A 15 14
FP40-ME FP50-ME
FPW50-ME
Working temperature range °C -40 ... 200 -50 ... 200
Temperature stability °C ±0,01 ±0,01
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20 -30
0.68 0.5 0.32 0.04
+20 0 -20 -30 -40
0.9 0.8 0.5 0.32 0.16
Refrigerant R404A R404A or R507
Overall dimensions (WxDxH) cm 37x46x69 42x49x70
Bath opening (WxL) cm 23x14 18x12
Bath depth cm 20 15
Filling volume from ... to liters 9 ... 16 5.5 … 8
Weight kg 48 55
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current draw (at 230 V) A 13 14
Current draw (at 230 V) CH A ------ 9 + 4
Mains power connection 230 V/60 Hz V/ Hz 207-253 / 50 207-253 / 60
Current draw (at 230 V) A 14 14
All measurements have been carried out at:
rated voltage and frequency ambient temperature: 20 °C
Technical changes without prior notification reserved.