9.6. MENU LIMITS
Menu level 1
>Li< = LIMITS – limitation of operating temperature range
Maximum and minimum setpoint
Restriction of the adjustable temperature range.
Level 2 Parameter level
Press the
key if parameter is to be retained.
C
orrection function for parameters or values (prior
to OK).
• Integer digits flash, set by pressing
+
• Decimal digits flash, set by pressing
+
• Integer digits flash, set by pressing
+
• Decimal digits flash, set by pressing
+
Factory settings::
200 °C
(Setpiont High)
-50 °C
(Setpoint Low)
The limitation of the operating temperature range effects the
temperature setting in the menu with the key .
Only setting of working temperatures which lie within the
determined limits is possible
This applies to settings in the MENU ,
(refer to page 37)
and for settings in the MENU
high temperature low temperature
(refer to page 42)
The temperature values are automatically deferred into the limit
range.
Setting range: -50.0 °C ... +200.0 °C