MA
51
>tYP< - ATC TYPE
• The parameter flashes, switch by pressing
and
A >1-point<, >2-point< or >3-point<
calibration can be carried out.
The value > tt1< is only indicated
In addition the measured temperature value
>Ct 1< is saved during the next step.
• Integer digits flash, set by pressing
+
• Decimal digits flash, set by pressing
+
If only a 1-point calibration is carried out, the following menu items
are not indicated anymore
The value > tt2< is only indicated
In addition the measured temperature value
>Ct 2< is saved during the next step.
• Integer digits flash, set by pressing
+
• Decimal digits flash, set by pressing
+
If only a 2-point calibration is carried out, the following menu items
are not indicated anymore
The value > tt3< is only indicated
In addition the measured temperature value
>Ct 3< is saved during the next step.
• Integer digits flash, set by pressing
+
• Decimal digits flash, set by pressing
+