9.2. MENU Configuration
Menu level 1
CFG = MENU CONFIGURATION
A RESET can be effected only in the > < mode.
Switch off circulator by pressing the key and call up the menu
CONFIGURATION.
Level 2 Parameter level
Press the key if a parameter is to be
retained. Correction function for parameters and
values (prior to OK).
or
>3SP< - Switch on and off the 3-setpoint mode
The parameter flashes, set by pressing
+
>YES< - This function can be used by pressing
the key. (refer to page 37)
or
>rt< - Switch on and off remote control
• The parameter flashes, set by pressing
+
For remote control refer to 59
Connect RS232 with PC
or
>Aut< - Switch on and off autostart
• The parameter flashes, set by pressing
+
AUTOSTART on
AUTOSTART off.
See WARNING page 47
>rSt< (RESET) - Use this to reset all values to
factory setting.
• Return to factory settings by pressing
During the message – init - all parameters are
reset to factory settings.