Safety notes for the user
26
The user alone is responsible for the handling of these substances!
The following questions shall help to recognize possible dangers and to reduce
the risks to a minimum.
• Are all tubes and electrical cables connected and installed?
Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.
• Do dangerous steams or gases arise when heating?
Is an exhaust needed when working?
• What to do when a dangerous substance was spilled on or in the unit?
Before starting to work, obtain information concerning the substance and
determine the method of decontamination.
Notice: Check the safety installations at least twice a year!
• Excess temperature protection according to IEC 61010-2-010.
With a screwdriver turn back the adjustable excess temperature protection
until the shut-down point (actual temperature).
• Low level protection according to IEC 61010-2-010.
To check the function of the float, it can be manually lowered with a
screwdriver for example.