Manual

Table Of Contents
MA
19
F33-MA F34-MA
Working temperature range
°C
-30 ... 200
-30 ... 150
Temperature stability
°C
±0,02
±0,02
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20 -30
0.5 0.32 0.12 0.03
+20 0 -20 -30
0.45 0.32 0.14 0.03
Refrigerant
R134a
R134a
Overall dimensions (WxDxH)
cm
36x46x69
38x58x62
Bath opening (WxL)
cm
23x14
24x30
Bath depth
cm
20
15
Filling volume from ... to
liters
12 ... 16
14 ... 20
Weight
kg
44
42
Mains power connection 230 V/50 Hz
V/ Hz
207-253 / 50
207-253 / 50
Current draw (at 230 V)
A
12
12
Current draw CH (at 230 V)
A
9 + 2
9 + 2
Mains power connection 230 V/60 Hz
V/ Hz
207-253 / 60
207-253 / 60
Current draw (at 230 V)
A
12
12
Current draw (at 208 V)
A
12
13
Mains power connection 115 V/60 Hz
V/ Hz
103-127 / 60
103-127 / 60
Current draw (at 115 V)
A
15
14
Mains power connection 100 V/50-60 Hz
V/ Hz
90-110/50-60Hz
-----
Current draw (at 100 V)
A
15
-----
FP35-MA
Working temperature range
°C
-35 ... 150
Temperature stability
°C
±0,02
Cooling capacity
Medium ethanol
°C
kW
+20 0 -20 -30
0.45 0.39 0.15 0.05
Refrigerant
R134a
Overall dimensions
(WxDxH)
cm
31x42x64
Bath opening (WxL)
cm
18x12
Bath depth
cm
5
Filling volume from ... to
liters
1,7 ... 2.5
Weight
kg
37
Mains power connection 230 V/50 Hz
V/ Hz
207-253 / 50
Current draw (at 230 V)
A
12
Current draw CH (at 230 V)
9 / 2
Mains power connection 115 V/60 Hz
V/ Hz
103-127 / 60
Current draw (at 115 V)
A
14
Mains power connection 100 V/50-60 Hz
V/ Hz
90-110/50-60Hz
Current draw (at 100 V)
A
14
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C Technical changes without prior notification reserved.