User Manual
HE
81
Command Parameter Response of circulator
out_par_12 xxx Tv control parameter of the cascade controller. 0 ... 999
out_par_13 xxx Maximum internal temperature of the cascade controller.
out_par_14 xxx Minimum internal temperature of the cascade controller.
out_par_15 xxx Band limit (upper) 0 ... 200 °C
out_par_16 xxx Band limit (lower) 0 ... 200 °C
out_hil_00 -xxx Set the desired maximum cooling power (0 % to 100 %).
Note: Enter the value with a preceding negative sign!
This adjustment only makes sense if FP refrigerating units are
used.
out_hil_01 xxx Set the desired maximum heating power (10 % to 100 %).
in commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of circulator
version none Number of software version (V X.xx)
status none Status message, error message (see page 83)
in_pv_00 none Actual bath temperature.
in_pv_01 none Heating power being used (%).
in_pv_02 none Temperature value registered by the external Pt100 sensor.
in_pv_03 none Temperature value registered by the safety sensor.
in_pv_04 none Setpoint temperature of the excess temperature protection
in_sp_00 none Working temperature „SETP 1“
in_sp_01 none Working temperature „SETP 2“
in_sp_02 none Working temperature „SETP 3“
in_sp_03 none High temperature warning limit „OverTemp“
in_sp_04 none Low temperature warning limit „SubTemp“
in_sp_05 none Setpoint temperature of the external programmer
(REG+E-PROG) .
in_sp_07 none 1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotation speed
after start.
in_sp_08 none Value of a flowrate sensor connected to the E-Prog input
in_par_00 none Temperature difference between working sensor and safety
sensor
in_par_01 none Te - Time constant of the external bath.
in_par_02 none Si - Internal slope
in_par_03 none Ti - Time constant of the internal bath.