User Manual

HE
65
BAND HIGH / LOW Band limitation
The band limitation is active during external control. Varied, practice-
oriented setting are feasible for heat-up and cool-down phases.
Setting range: 0 °C ... 200 °C
BAND HIGH and BAND LOW allow for the limitation of the difference
between the temperature
s in the internal bath and the external system to
any maximum value for the heat-up and cool-down phase.
During the heat-up phase this difference value is always added to the
actual external temperature. During the cool-down phase the difference
value is subtracted.
25
50
10
20
30
40
50
60
10
20
30
40
50
60
10
[min]
75
100
150
200
°C
t
external system
Internal bath
BAND HIGH
BAND LOW
Sense of a band limitation:
Protection of objects and samples
by gentle temperature control
Protection of e.g. glass reactors
from thermal shock.