Owner manual

Operator responsibility – Safety recommendations
10
2.4. Technical specifications
Ultra-Low Refrigerated Circulator
FP51-SL
Working temperature range °C -51 ... 200
Temperature stability °C ±0,05
Cooling capacity °C +20 0 -20 -40
Medium: ethanol kW 2.0 1.5 1.0 0.26
Cooling compressor 1-stage
Refrigerant R507
Overall dimensions (WxDxH) cm 46x55x89
Bath opening (WxL) cm 18x12
Bath depth cm 20
Filling volume liters 7,5 ... 12
Weight kg 90
Ambient temperature °C 5 ... 40
Mains power connection 400 V/3P/50 Hz V/ Hz 360-440V/3PNPE/50Hz
Current draw (at 400 V) A 14
Mains power connection 230 V/3P/60 Hz V/ Hz 207-253V/3PPE/60Hz
Current draw (at 230 V) A 14
Mains power connection 230 V/3P/50 Hz V/ Hz 207-253V/3PPE/50Hz
Current draw (at 230 V) A 14
SL
Temperature selection digital
via keypad indication on LCD DIALOG-DISPLAY (°C/°F)
remote control via PC indication on monitor
Temperature indication VFD COMFORT-DISPLAY (°C/°F)
Resolution °C 0.01
ATC3 INT / EXT °C ±3 / ±9
Temperature control ICC - Intelligent Cascade Control
Heater wattage (at 230 V) kW 3.0
Electronically adj. pump capacity stages 1 ... 4
Flow rate at 0 bar l/min 22 ... 26
Max. pressure at 0 liters bar 0.7
Max. suction at 0 liters bar 0.4
Electrical connections: see page 11
Ambient temperature °C 5 ... 40
Mains power connection 230 V/50-60 Hz V/ Hz 190 ... 253 / 50-60
Current draw (at 208 V / 230 V) A 12 / 13
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C Technical changes without prior notification reserved.