12.3. List of commands
out commands: Setting temperature values or parameters.
Command Parameter Response of circulator
out_mode_01 0 Use working temperature > SETPNT 1<
out_mode_01 1 Use working temperature > SETPNT 2<
out_mode_01 2 Use working temperature > SETPNT 3<
out_mode_02 0 Selftuning „off“.
Temperature control by using the stored parameters.
out_mode_02 1 Selftuning „once“
Single selftuning of controlled system after the next start.
out_mode_02 2 Selftuning „always“
Continual selftuning of controlled system whenever a new
setpoint is to be reached.
out_mode_03 0 Set external programmer input to voltage.
Voltage 0 V ... 10 V
out_mode_03 1 Set external programmer input to current.
Current 0 mA ... 20 mA
out_mode_04 0 Temperature control of internal bath.
out_mode_04 1 External control with Pt100 sensor.
out_mode_05 0 Stop the unit = R –OFF-.
out_mode_05 1 Start the unit.
out_mode_08 0 Set the control dynamics - aperiodic
out_mode_08 1 Set the control dynamics - standard
out_sp_00 xxx.xx Set working temperature. „SETPNT 1“
out_sp_01 xxx.xx Set working temperature. „SETPNT 2“
out_sp_02 xxx.xx Set working temperature. „SETPNT 3“
out_sp_03 xxx.xx Set high temperature warning limit „OverTemp“
out_sp_04 xxx.xx Set low temperature warning limit „SubTemp“
out_sp_06 xxx.xx Set manipulated variable for the heater via serial interface
-99.99 ... +100 [%]
out_sp_07 x Set the pump pressure stage. (1 ... 4)
out_par_04 x.x CoSpeed for external control 0 ... 5.0
out_par_06 xxx Xp control parameter of the internal controller. 0.1 ... 99.9
out_par_07 xxx Tn control parameter of the internal controller. 3 ... 9999
out_par_08 xxx Tv control parameter of the internal controller. 0 ... 999
out_par_09 xxx Xp control parameter of the cascade controller. 0.1 ... 99.9
out_par_10 xxx Proportional portion of the cascade controller. 1 ... 99.9
out_par_11 xxx Tn control parameter of the cascade controller. 3 ... 9999