Owner's manual
30
The out commands are valid only in remote control mode.
Examples:
Command to set the working temperature to15,5 °C:
out_sp_00
15.5
Command to ask for the working temperature
in_sp_00
Response from the recirculating cooler:
15.5 LF
11.3. List of commands
out commands: Setting parameters or temperature values.
Command Parameter Response of recirculating cooler
out_mode_05 0 Stop the unit = R –OFF-.
out_mode_05 1
Start the unit.
out_sp_00 xxx.xx
Set working temperature
in commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of recirculating cooler
version none Number of software version (V X.xx)
status none
Status message, error message (see page 30 )
in_pv_00 none Actual bath temperature.
in_sp_00 none Working temperature
in_mode_05 none
Recirculating cooler in Stop/Start condition:
0 = Stop
1 = Start
11.4. Status messages
Status messages Description
00 MANUAL STOP
Recirculating cooler in „OFF“ state.
01 MANUAL START
Recirculating cooler in keypad control mode.
02 REMOTE STOP
Recirculating cooler in „r OFF“ state.
03 REMOTE START
Recirculating cooler in remote control mode.