Instruction Manual
Operator responsibility – Safety recommendations
10
2.4. Technical specifications
Calibration Baths FK30-SL FK31-SL
Working temperature range °C -30 ... 200 -30 ... 200
Temperature stability
(1
°C ±0.005 ±0.005
Temperature uniformity °C <±0.01 <±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20
0.46 0.34 0.15
+20 0 -20
0.46 0.34 0.15
Refrigerant R134a R134a
Overall dimensions (WxDxH) cm 32x45x79 32x45x91
Bath opening (WxL) cm
12 12
Bath depth cm 17 31
Filling volume Liter 14 24
Weight kg 48 51
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current draw (at 230 V)
Current draw (at 230 V) <GB>
A
A
A
14
10 (Circulator)
4 (Cooling machine)
14
10 (Circulator)
4 (Cooling machine)
Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60 103-127 / 60
Current draw (at 115 V) A
A
10 (Circulator)
6 (Cooling machine)
10 (Circulator)
6 (Cooling machine)
(1
±0.02 °C@+200 °C
Circulator (SL-C) SL
Temperature selection digital
via keypad indication on LCD DIALOG-DISPLAY (°C/°F)
remote control via PC indication on monitor
Temperature indication VFD COMFORT-DISPLAY (°C/°F)
Resolution °C 0.01
ATC3 INT / EXT °C ±3 / ±9
Temperature control ICC - Intelligent Cascade Control
Heater wattage (at 230 V) kW 2.0
Heater wattage (at 115 V) kW 1.0
Electronically adj. pump capacity stages 1 ... 4
Flow rate at 0 bar l/min 22 ... 26
Max. pressure at 0 liters bar 0.7
Max. suction at 0 liters bar 0.4
Ambient temperature °C 5 ... 40
Mains power connection 230 V/50-60 Hz V/ Hz
190 ... 253 / 50-60
Current draw (at 208 V / 230 V) A
12 / 13
Mains power connection 115 V/50-60 Hz V/ Hz
90 ... 127 / 50-60
Current draw (at 100 V / 115 V) A
9/10
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C Technical changes without prior notification reserved.