Owner manual
Operator responsibility – Safety instructions
8
2.4. Technical specifications
FT200 FD200
Temperature range °C -20 … +30 10 … +30
Cooling capacity
(medium ethanol)
°C
kW
+20 0 -30
0.25 0.15 0.04
+20 +10
0.22 0.18
Refrigerant R134a R134a
Recommanded flow rate l/min ----- 2 - 3
Freezing protection °C ----- 10
Immerson probe (Lxdia.) cm 9x4 -----
Connection tubing (L) cm 120 -----
Dimensions (WxLxH) cm 18x27x39 18x27x39
Weight kg 18 16
Ambient temperature °C 5 ... 35 5 ... 35
Mains power connection V/Hz 190-253 /50 230/50
Current draw (at 230 V) A 1.5 1.5
Mains power connection V/Hz 103-127 / 60 115 / 60
Current draw (at 115 V) A 3.0 3.0
FT400 FT900
Temperature range °C -40 ... +30 -90 ... +30
Cooling capacity
(medium ethanol)
°C
kW
+20 +10 -20 -40
0.45 0.36 0.14 0.03
+20 +10 -40 -80
0.3 0.27 0.2 0.7
Cooling compressor 1-stage 2- stage
Refrigerant R404A R404A/R23
Immerson probe (Lxdia.) cm 12x5 65x1.5 (flexible)
Connection tubing (L) cm 120 160
Dimensions (WxLxH) cm 20x30x43 38x55x60
Weight kg 24 50
Ambient temperature °C 5 ... 35 5 ... 35
Mains power connection V/Hz 190-253 /50 190-253 / 50-60
Current draw (at 230 V) A 4 5.5
Mains power connection V/Hz 103-127 / 60 103-127 / 60
Current draw (at 115 V) A 3.5 7.0
Note:
All measurements have been carried out at:
rated voltage and frequency; ambient temperature 20 °C;