6.4. PID control parameters
For internal and external control two separate parameter
sets are available.
The PID control parameters can be adapted to the
requirements of the controlled member.
The values are preserved after switching off the
recirculating cooler.
SET
ACTUAL
SET
The control parameters are indicated by operating the
key (5.1).
Indications in case of internal control INT:
Parameter Setting range
CP 1 Xp (example 4.0 K). 0.1 ... 100 K
CP 2 Tn (example 160 s). 1 ... 9999 s
CP 3 Tv (example 20 s). 0 ... 500 s
Each indicated control parameter can be optimized
manually .
Setting
:
Operate the key as often until the desired control
parameter is indicated. Example: CP3
Use the cursor keys to move left or right on
the display until the numeral you wish to change is
blinking.
Use the increase/decrease arrows to
change the selected numeral (-, 0, 1, 2, 3, ... 9).
Press enter to store the selected value