ACTUAL
SET
Ad
¯
P - Calibration of programmer input:
Highest value = 80 °C
ACTUAL
SET
Ad
P - Calibration of programmer input:
Lowest value = -20 °C
ACTUAL
SET
Calibration procedure:
In the configuration level, set the programmer input
type to allow control via an external programmer.
0 = Voltage 0 to 10 V
1 = Current 0 to 24 mA
Connect an external programmer to the connector (O).
Set the external programmer to the highest
temperature value.
ACTUAL
SET
In the calibration level, select the item Ad
¯
P and set
the highest temperature value
(example: 80 °C).
Press the setpoint key (E).
Follow the instructions under section 6.3.1.
page 27
Do not alter any settings on either of the units as long
as the display "Ad¯P" is blinking.