Manual

Operator responsibility – Safety instructions
10
(with S-pump) FC1600S FCW1600S
Working temperature range °C -15 ... 80 -15 ... 80
Cooling capacity °C 20 10 5 -10
(water-glycol) kW 1.55 1.15 0.9 0.36
Refrigerant R134a
Heater capacity kW 1.2
Pump capacity:
Pressure max. bar 1.2
Flow rate max. l/min 22 / 15
with tubing connections mm dia. 13.5 / 9.5
Pump connections M16x1
Noise level, 1 m distance dBA 57 53
Filling volume l 8 ... 11
Dimensions (WxLxH) cm 46 x 61 x 49
Ambient temperature °C 5 ... 40
Shipping weight kg 66 68
Mains power connection V/Hz 208-230/60 or 230/50
Total power consumption W 2900
(with Ju-pump) FC600 -T
Working temperature range °C -20 ... 80
Cooling capacity °C 20 10 5 -10
(water-glycol) kW 0.6 0.47 0.4 0.21
Refrigerant R134a
Heater capacity kW 1.2
Pump capacity:
Pressure max. bar 0.5
Flow rate max. Lpm 20 / 14
with tubing connections mm dia 13,5/9,5
Pump connections M16x1
Noise level, 1 m distance dBA 51
Filling volume l 6 ... 8
Dimensions (WxLxH) cm 46 x 61 x 49
Ambient temperature °C 5 ... 40
Shipping weight kg 48
Mains power connection V/Hz 208-230/60 or 230/50
Total power consumption W 2000
All data have been determined at rated voltage and frequency
ambient temperature: 20 °C bath liquid: water-glycol