9. Set MENU Program to >Mode Forcier<
See page 32
10. Start the forcing test
As factory-setting the „Forcing test 60 °C“ is stored
Measuring cycle:RUNS 1 flow (48 h)
Warm phase W-TEMP 60 °C
Duration W-TIME 24 h
Cool phase C-TEMP 0 °C
Duration C-TIME 24 h
Setting of other values see page 32 FORCING-Test Program
Start-Menu
Press the key if a parameter is to be retained.
Correction function for parameters or values (prior to OK)
Menu level 1
> RUNS < Number of repetitions of the warm/cool phase 1 ... 10
> GO < Start the forcing test (NOW)
Level 2 Parameter level Level 2