User guide

Menu functions
32
9.2. MENU PROGRAM – creation, administration
Program Mode
Set this menu item first!
In the >MENU PROGRAM< you can choose between the „Standard Program“
and the „Forcing Test-Program“.
This setting determines which submenu items are available in >PROGRAM
EDIT< and >PROGRAM DELETE<.
Menu level 1
Press key, if a parameter is to be retained.
Correction function for parameters or values (prior to OK)
Level 2 Level 3
Set program mode with and
FORCING-Test Program page 32
STANDARD-Program page 34
9.2.1. FORCING-Test Program
>EDIT< PROGRAM – creation, administration
> W-TEMP< Temperature of the warm phase
Adjustable from –40 °C up to +80 °C
>W-TIME < Duration of the warm phase
Adjustable max. 999 h 59 minutes
> C-TEMP< Temperature of the cool phase
Adjustable from –40 °C up to +80 °C
>C-TIME < Duration of the cool phase
Adjustable max. 999 h 59 minutes
> DELETE< without function when forcing test program is set
>MODE< Program Mode