Instruction Manual

EH
9
2.4. Technical specifications
F12-EH F25-EH
Working temperature range °C -20 ... 150 -28 ... 150
Temperature stability °C ±0.03 ±0.03
Temperature selection digital digital
Temperature indication LED LED
Resolution °C 0.1 0.1
ATC - Absolute Temperature Calibration 1-point 1-point
Temperature control PID1 PID1
Heater wattage (at 230 V) kW 2.0 2.0
Heater wattage (at 115 V) kW 1.0 1.0
Cooling capacity °C +20 0 -20 +20 0 -20
Medium ethanol kW 0.16 0.1 0.02 0.26 0.2 0.06
Refrigerant R134a R134a
Circulating pump:
discharge, max.at 0 bar l/min 15 15
pressure, max. at 0 l bar 0,35 0,35
Electrical connections:
External alarm device Vdc/mA 24-0 / max. 25 24-0 / max. 25
Overall dimensions (WxDxH) cm 20x36x56 23x42x61
Bath opening (WxL) cm 13x15 12x14
Bath depth cm 13 14
Filling volume from ... to liters 3 ... 4.5 3 ... 4.5
Weight kg 22 30
Ambient temperature °C 5 ... 40 5 ... 40
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current input (at 230 V) A 11 12
Current input (at 230 V) CH A <9 + 1> <9 + 2>
Mains power connection 230 V/60 Hz V/ Hz 207-253 / 60 207-253 / 60
Current input (at 230 V) A 11 12
Mains power connection 115 V/60 Hz V/ Hz 103-127 / 60 103-127 / 60
Current input (at 115 V) A 12 13
Mains power connection 100 V/50-60 Hz V/ Hz 90-110V/50-60Hz 90-110V/50-60Hz
Current input (at 100 V) A 15 13
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C Technical changes without prior notification reserved.