Owner manual

Remote control
72
Command Parameter Response of Cryo-Compact Circulator
in_par_11 none Tn control parameter of the cascade controller.
in_par_12 none Tv control parameter of the cascade controller.
in_par_13 none Adjusted maximum internal temperature of the cascade controller.
in_par_14 none Adjusted minimum internal temperature of the cascade controller.
in_par_15 none Band limit (upper)
in_par_16 none Band limit (lower)
in_mode_01 none Selected setpoint:
0 = Setpoint 1
1 = Setpoint 2
2 = Setpoint 3
in_mode_02 none Selftuning type:
0 = Selftuning „off“
1 = Selftuning „once“
2 = Selftuning „alway“
in_mode_03 none Type of the external programmer input:
0 = Voltage 0 V to 10 V
1 = Current 0 mA to 20 mA
in_mode_04 none
Internal/external temperature control:
0 = Temperature control with internal sensor.
1 = Temperature control with external Pt100 sensor.
in_mode_05 none
Cryo-Compact Circulator in Stop/Start condition:
0 = Stop
1 = Start
in_mode_08 none
Adjusted control dynamics
0 = aperiodic
1 = standard
in_hil_00 none Max. cooling power (%).
in_hil_01 none Max. heating power (%).
11.4. Status messages
Status messages Description
00 MANUAL STOP
Cryo-Compact Circulator in „OFF“ state.
01 MANUAL START
Cryo-Compact Circulator in keypad control mode.
02 REMOTE STOP
Cryo-Compact Circulator in „r OFF“ state.
03 REMOTE START
Cryo-Compact Circulator in remote control mode.