Owner manual

Cryo-Compact Circulators
71
in commands: Asking for parameters or temperature values to be displayed.
Command Parameter Response of Cryo-Compact Circulator
version None Number of software version (V X.xx)
status none Status message, error message (see page 72)
in_pv_00 none Actual bath temperature.
in_pv_01 none Heating power being used (%).
in_pv_02 none Temperature value registered by the external Pt100 sensor.
in_pv_03 none Temperature value registered by the safety sensor.
in_pv_04 none Setpoint temperature („SafeTemp“) of the excess temperature
protection
in_sp_00 none Working temperature „Setpoint 1“
in_sp_01 none Working temperature „Setpoint 2“
in_sp_02 none Working temperature „Setpoint 3“
in_sp_03 none High temperature warning limit „OverTemp“
in_sp_04 none Low temperature warning limit „SubTemp“
in_sp_05 none Setpoint temperature of the external programmer
(REG+E-PROG) .
in_sp_06 none Temperature indicatio in °C or °F
in_sp_07 none 1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotation speed
after start.
in_sp_08 none Value of a flowrate sensor connected to the E-Prog input
in_par_00 none
Temperature difference between working sensor and safety sensor
in_par_01 none Te - Time constant of the external bath.
in_par_02 none Si - Internal slope
in_par_03 none Ti - Time constant of the internal bath.
in_par_04 none CoSpeed - Band limit (max. difference between the temperatures
in the internal bath and external system).
in_par_05 none Factor pk/ph0:
Ratio of max. cooling capacity versus max. heating capacity
in_par_06 none Xp control parameter of the internal controller.
in_par_07 none Tn control parameter of the internal controller.
in_par_08 none Tv control parameter of the internal controller.
in_par_09 none Xp control parameter of the cascade controller.
in_par_10 none Proportional portion of the cascade controller.