Owner manual

Menu functions
36
8.2. Control
The Cryo-Compact Circulator is conceived for internal and external
temperature control. Switching is carried out in this submenu. Depending on
what is set, only the respective set of parameters is indicated.
M
ain menu
Profile Start
Int. Programmer
>Control
C
ontrol
>
ControlType:
Selftuning : once /always/off
Dynamic : aperio /normal
Xp : 1.0K
Tn : 100s
Tv : 5s
i
ntern
C
ontrol
>
ControlType: extern
Selftuning: once /always/off
CoSpeed : 0.0
Xp : 0.7K
Tn : 720s
Tv : 55s
Xpu : 5.0K
1
ESC
ESC
internal
external
Example: 0.8
1. Use the keys to select the menu item.
2. With the keys select the parameter (parameter blinks)
or
use the numeral keypad to enter the numbers (number segment blinks).
3. Press enter to store the new parameter or value.
-OFF-
S xx.xx
E xx.xx
ControlType: Can only be adjusted in the -OFF- condition
Possible parameters:
intern internal temperature control
extern external temperature control with external Pt100 sensor
Pt100
IMPORTANT:
Additional measures for external temperature control.
For external control and temperature measurement an external Pt100
sensor must be connected to the socket (7) on the rear of the Cryo-
Compact Circulator .