User Manual

Operator responsibility – Safety instructions
8
2.4. Technical specifications
AWC100
Working temperature range °C +20 ... +40
Cooling capacity varying with the temp. Difference
between return line temp. and ambient temperature
stage 0
stage 1
°C
W
W
20 15 10 5
400 320 220 120
550 440 300 180
Circulating pump
flow rate, max. l/min 2.9
pressure, max. bar 0,2
Noise level, 1 m distance max dB(A) 55
Filling volume (internal storage tank) l 0.85
Overall dimensions (WxLxH) cm 20x34x30
Ambient temperature °C 5 ... 40
Weight kg 11
Mains power connection 230 V/50-60 Hz V/Hz 207-253V/50-60
Current draw (at 230 V) A 1
Mains power connection 115 V/60 Hz V/Hz 103-127V/60
Current draw (at 115 V) A 1
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C .
Environmental conditions according to IEC 61 010-1:
Use only indoor.
Altitude up to 2000 m - normal zero.
Ambient temperature: +5 ... +40 °C
Air humidity:
Max. rel. humidity 80 % for temperatures up to +31 °C,
linear decrease down to 50 % relative humidity at a temperature of +40 °C
Max. mains fluctuations of ±10 % are permissible.
The unit corresponds to Class I
Overvoltage category II
Pollution degree 2
Caution:
The unit is not for use in explosive atmosphere