Owner's Manual

Important Testing Guidelines
8
109
1.
2. Sit in a calm environment for at least 5 minutes prior to testing.
3. Do not stand while testing. Sit in a relaxed position while keeping your arm level with
your heart.
4. Avoid speaking or moving body parts while testing.
5. While testing, avoid strong electromagnetic interference such as microwave ovens and
cell phones.
6. Wait 3 minutes or longer before re-testing.
7. Try to measure your blood pressure at the same time each day for consistency.
8. Test comparisons should only be made when monitor is used on the same arm, in the
same position, and at the same time of day.
9. This blood pressure monitor is not recommended for people with severe arrhythmia.
:
1. The position of the subject, his or her physiologic condition;
2. The performance and accuracy of the device;
3. Cuff size: too small cuff (bladder) will produce a higher blood pressure value than usual,
too big cuff (bladder) will produce a lower blood pressure value;
4. Measuring position does not keep level with your heart;
5. Speaking or moving body parts while testing;
6. Not relaxing for about 5 minutes before taking the measurement.
Avoid eating, exercising, and bathing for 30 minutes prior to testing.
10. Do not use this blood pressure monitor if the device is damaged.
Any blood pressure recording can be affected by the following factors
Quick Start
1. Install batteries
2. Insert cuff air plug into the behind side of monitor unit. (See Figure B)
. (See Figure A )
3. Remove thick clothing from the arm area.
4. Rest for several minutes prior to testing. Sit down in a quietplace comfortably ,
back and arm support on a desk or table ,withyour legs uncrossed ,your arm
resting on a firm and your feet flat on the floor. (See Figure C )
5. Apply cuff to your left arm and middle of the cuff at the level of your heart. Bottom
of cuff should be placed approximately 1-2cm (1/2'') above elbow joint. (See Figures D&E)
1-
2cm
'
'
(0
.4
-
0.
8 )
Unit Operation
Battery Installation
Slide battery cover off as indicated by arrow.
Install 3 new AAA alkaline batteries according to polarity.
Close battery cover.
Figure B
Figure A
Figure C Figure D
6. Press " START/STOP " Button to start testing.
Unit Illustration
7
Contents
Note:Power supply is specified as part of ME EQUIPMENT.
Display
Group1
Group2
Systolic
Blood
Pressure
Diastolic
Blood
Pressure
Pulse Rate
Irregular Heartbeat Indicator
Time/Date
Heart Rate Indicator
WHO Blood Pressure
Classification Indicator
Last 3 Tests
Average
Battery indicator
Arm Shake Indicator
Cuff Loose Indicator
IHB
1
2
3
Bluetooth
SMART
bluetooth indicator
3.Storage Case
2.Owner's Manual
1.Monitor Unit
5.2MOPPMedical AC Adapter(DC5.0 V,1000mA)
(recommended, not provided)
Arm-type F ully Automatic
Digital Bl ood P ressure Monito r
臂 式
Mode lDB P-629 3B
USB Adapter JackUSB Adapter Jack
This blood pressure provides external power interface, can use the 2MOPP medical power
adapter. The power supply port of the device is USB-C port, and the port applicable parameters
is voltage 5v and current 1000mA.Do not use other types of AC adapters.otherwise, the device
may be damaged.