Owner's Manual
45
Warranty
The Blood Pressure Monitor is guaranteed for 2-year from the
date of purchase. If the Blood Pressure Monitor does not function
properly due to defective components or poor workmanship, we
will repair or replace it freely. The warranty does not cover
damages to your Blood Pressue Monitor due to improper handling.
Please contact local retailer for details.
44
Specifications
Electromagnetic Compatibility Information
46
The device satisfies the EMC requirements of the international standard IEC
60601-1-2. The requirements are satisfied under the conditions described in the
table below. The device is an electrical medical product and is subject to special
precautionary measures with regard to EMC which must be published in the
instructions for use. Portable and mobile HF communications equipment can affect
the device. Use of the unit in conjunction with non-approved accessories can affect
the device negatively and alter the electromagnetic compatibility. The device
should not be used directly adjacent to or between other electrical equipment.
Table 1
Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Electromagnetic environment
-guidance
ComplianceEmissions test
The device uses RF energy only for
its internal function. Therefore, its
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
Group 1,
ClassB
Radiated emission CISPR 11
Conducted emission CISPR 11
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
N/A
N/A
43
Specifications
42
Specifications
41
Specifications
Pressurization
Memory
Function
2x60 Memories in Tow Groups with Date and Time
Irregular Heartbeat Detection
WHO Classification Indicator
Last 3 Results Average
Low Battery Detection
Automatic Pressurization
Automatic Power-Off
Product Description
Model
Display
Measurement Method
Oscillometric Method
Wrist-Type Fully Automatic
Digital Blood Pressure Monitor
DBP-8276H
Measurement Range
Pulse
Systolic Pressure
Pulse
60mmHg~260mmHg
30 ~ 180 Beats/Minute
±5%
Diastolic Pressure
30mmHg~200mmHg
Pressure
0mmHg~299mmHg
Pressure
±3mmHg
LCD Digital Display
Size:34mm×34.6mm (1.34" x 1.36")
Power Source
Battery run time
Unit Weight
Cuff Circumference
lithium battery(3.7v 180mAh)
500 measurements
Fits wrist circumference 13.5-21.5 cm(5.3"-8.5")
Humidity
Pressure
Operating Environment
Temperature
Storage Environment
Humidity
Temperature
Function
Backlight
Buzzer
15%~93%RH
800hPa~1060hPa
10℃ ~ 40℃ (50 ~104 )℉ ℉
-25℃~55℃ (-13 ~131 )℉ ℉
≤93% RH
Approx. 122g (4.30oz.) ( Excluding Battery)
Approx. 62mm×55.2mm×19mm(L x W x H)
(2.44 " x 2.17"x 0.75")
Unit Dimensions
1 hours
Charge time
Transport Environment
-25℃~55℃ (-13℉~131℉)
Humidity
Temperature
≤93% RH
Classification
Internal Powered Equipment Type BF
IP 22
Ingress Protection Rating
Battery Storage
Temperature
-25℃~55℃ (-13℉~131℉)
Battery Shelf life
60 months
Version
Modulation Type
GFSK
5.0.1 BT Signal mode
Operation
frequency
2.4GHz(2400~2483.5MHz)
Antenna gain
0.5 dBi
Bluetooth
This Blood Pressure Monitor complies with the European regulations
and bears the CE mark“CE 0197”.This blood pressure monitor also
complies with mainly following standards
(included but not limited):
Safety standard:
EN 60601-1 Medical electrical equipment part 1: General
requirements for safety
EMC standard:
EN 60601-1-2 Medical Electrical Equipment -- Part 1-2: General
Requirements For Basic
Safety And Essential Performance -- Collateral Standard:
Electromagnetic Disturbances –
Requirements And Tests.
Performance standards:
IEC80601-2-30, Medical electrical equipment – Part 2-30: Particular
requirements for the basic safety and essential performance of
automated non-invasive sphygmomanometers.
EN 1060-3 Non-invasive sphygmomanometers - Supplementary
requirements forelectromechanical blood pressure measuring
systems.
ISO 81060-2, non-invasive sphygmomanometers - part 2: clinical
validation of automated measurement type.
Specifications are subject to change without notice.
Bandwidth
2.0 MHz