4. Cuff Cleaning and Disinfection:
A) Spread the cuff (skin-contact surface) upwards onto a clean table. Use a damp clean cloth
(water-based) to wipe the skin-contact surface with a force.
B) Soak the cloth clean with drinking water and wring it dry. Repeat A) with the damp cloth
(water-based) for 3 times.
C) Apply 70%-80% alcohol to a new cloth (or 75% alcohol cotton-ball), use it to
wipe the skin-contact surface with a force. Then soak the cloth with the alcohol again
(or change a new 75% alcohol cotton-ball), repeat the disinfection procedure for 3 times.
D)When the disinfection towards the skin-contactsurface is finished,wipe thenon- skin
contact surface with a cloth (alcohol-based) or alcohol cotton-ball thoroughly for
3 times.
E)Leave the cuff naturally dry,then it is ready for reuse.
Notice: Do not soak in water or splash water on it.
5. Do not use petrol, thinners or similar solvents.
6. Remove batteries when not in operation for an extended period of time.
7. Do not disassemble product.
8. It is recommended the performance should be checked every 2 years.
9. Expected service life: Approximately three years at 10 tests per day.
10.No service and maintenance while it is in use and maintenance only be performed by service
personnel. Service and maintenance require parts, repair, technical support will be provided.
Maintenance
27
Product Description
Model
Display
Measurement Method
Oscillometric Method
Arm-type Fully Automatic Blood Pressure Monitor
DBP-6281B
Pressurization
Memory
Function
Power Source
Battery Life
Unit Weight
Unit Dimensions
Cuff Circumference
2x60 Memories in Two Groups with Date and Time
Irregular Heartbeat Detection
WHO Classification Indicator
Low Battery Detection
Automatic Power-Off
Voice
3 AAA batteries or Medical AC Adapter(DC5.0V, 1000mA)
(recommended, not provided)
Approximately 2 months at 3 tests per day
Approx.341g (12.03 oz.) (excluding battery)
Approx.155.5mm X 108.5mm X 75mm( 6.12" x 4.27" x 2.95") (L x W x H)
Operating Environment
Automatic Pressurization
LCD Digital Display Size:87.5mm X 85mm( 3.44" x 3.34")
Humidity
Temperature
10 ~ 40 (50 ~104 )℃ ℃ ℉ ℉
15% ~ 93% RH
Storage Environment
Humidity
Classification:
Ingress Protection
rating:
Battery Shelf life:
Battery Storage
Temperature:
Temperature:
Internal Powered Equipment,Type BF . Cuff is the Applied Part
IP 20,Indoor Used Only
60 months
-25℃~55℃ (-13℉~131℉)
-25 ~55 (-13 ~131 )℃ ℃ ℉ ℉
Measurement Range
Pulse
Systolic Pressure
Pulse
60mmHg~26 0 mmHg
30 ~ 180 Beats/Minute
±5%
Diastolic Pressure
30mmHg~200 mmHg
Pressure
0mmHg~299mmHg
Pressure
±3mmHg
Backlight
≤93% RH
Medium cuff: Fits arm circumference 22-36 cm
Operating Environment
Pressure
800hPa~1060hPa
Bluetooth
Last 3 Tests Average
1. Avoid dropping, slamming, or throwing the unit.
2. Avoid extreme temperatures. Do not expose unit directly under sunshine.
3. When cleaning the unit, use a soft fabric and lightly wipe with mild detergent.
Use a damp cloth to remove dirt and excess detergent.
This Blood Pressure Monitor complies with the European regulations and bears the
CE mark“CE 0197”.This blood pressure monitor also complies with mainly
following standards
(included but not limited):
Safety standard:
EN 60601-1 Medical electrical equipment part 1: General requirements for safety
EMC standard:
EN 60601-1-2 Medical Electrical Equipment -- Part 1-2: General Requirements For Basic
Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances –
Requirements And Tests.
Performance standards:
IEC81060-2-30, Medical electrical equipment – Part 2-30: Particular requirements for the
basic safety and essential performance of automated non-invasive sphygmomanometers.
EN 1060-3 Non-invasive sphygmomanometers - Supplementary requirements for
electromechanical blood pressure measuring systems.
EN 1060-4 Non-invasive sphygmomanometers - Test procedures to determine the overall
system accuracy of automated non-invasive sphygmomanometers.
ISO 81060-2, non-invasive sphygmomanometers - part 2: clinical validation of automated
measurement type.
Specifications are subject to change without notice.
Bluetooth
Version
Modulation Type
GFSK
5.0.1 BT Signal mode
Operation frequency
2.4GHz(2400~ 2483.5MHz)
Antenna gain
0.5 dBi
Bandwidth
2.0 MHz