User Manual
www.joinwit.com
19
Image 5-38
(5)Close Software
Click the “Exit”button is OK.
vi、Troubleshooting
Fault Situations
Possible Cause
Solutions
Weak or no LCD display after
the instrument turning on
① Low battery
① change battery
Unstable source output power
① The internal source connection
port is severely polluted
② If the proper patch cord port is
connected with the source port
① Using the special cotton
swab to clean the port
② Using the correct patch
cord port
Large optical power meter
test error
① The internal source detecting
port is severely polluted
① Using the special cotton
swabs to clean the port
The instrument can’t be
turned on after the USB
adapter connection
① Mini USB cable is damaged
② USB adapter is damaged
① Change Mini USB cable
② Change to 5V USB adapter
Others
please contact the company
after-sales or technicians to
solve
vii、Care and Maintenance
1. Keep the sensor port clean from pollution; not use dirty and non-standard
fiber-optic adapter connector; not to be inserted to the badly-polished port. Otherwise
the sensor port can be damaged and the whole system performance will be affected.
2. Insist on the use of one fiber-optic adapter as possible.
3. Once the instrument is not used, cover it with dust cap to keep the port clean