Operator's Manual
Table Of Contents
- Cover page
- Front Matter
- Table of Contents
- 1 — Introduction
- 2 — Warnings, Cautions, and Other Residual Risks
- 3 — System Components
- 4 — Surgery Start Up
- 5 — Surgeon Settings
- 6 — Program Settings
- 7 — System Configuration
- Setting the Maximum IV Pole Height
- Testing the IV Pole
- Event Log
- Wireless Foot Pedal Setup
- VERITAS Remote Control Setup
- System Versions
- System Self Test
- Language Selection
- Surgical Media Center (SMC) (Optional)
- Set System Date and Time
- Set Units of Measure for Vacuum
- Service Interval
- Import/Export Database
- Backup All
- Print to File
- Restore Database
- Restore All
- Delete a Backup Database
- 8 — Symbol Definitions
- 9 — Care and Cleaning
- 10 — Error Messages, Troubleshooting, and Diagnostics
- 11 — Warranty and Maintenance
- 12 — Specifications
- Physical Specifications
- Environmental Specifications
- Electrical Specifications
- Electromagnetic Compatibility
- RF Wireless Technology
- Bluetooth Characteristics
- Phacoemulsification Specifications
- Irrigation and Aspiration Specifications
- Vitrectomy Specifications
- Diathermy Specifications
- Diathermy Power Graphs
- Diathermy Power versus Load Impedance
- Phaco Power Graphs
- Phaco Tip Velocity
- Bottle Height Pressure
- 13 — Accessories and Parts Reordering
- Index

WHITESTAR Technology Z370584 Rev. D
6-22
Program Settings
Table 6.7 – WHITESTAR Technology Parameter Settings
Figure 6-12 — Duty Cycle
Setting Pulse On Time (ms) Pulse Off Time (ms) Duty Cycle Pulse Rate (pps)
6/12 6 12 33% 55
4/8 4 8 33% 83
6/4 6 4 60% 100
6/8 6 8 43% 71
8/4 8 4 67% 83
4/24 4 24 14% 35
6/28 6 28 18% 29
6/24 6 24 20% 33
8/24 8 24 25% 31
6/18 6 18 25% 41